<?xml version="1.0" encoding="UTF-8"?><Articles><Article><id>36</id><JournalTitle>ANALYTICAL METHOD VALIDATION OF STABILITYINDICATING HPLC METHOD FOR DETERMINATION OF ASSAY OF MEMANTINE HYDROCHLORIDE TABLETS</JournalTitle><Abstract>This paper deals with the development and validation of stability indicating high performance liquid chromatographic
method for the quantitative determination of Memantine hydrochloride. Study of an analytical method for the determination of
Memantine hydrochloride drug and also validate the method as per the ICH guideline and required acceptance criteria. The
method was developed by using Hypersil BDS C8 column (250Ã—4.6 mm id, 5 Âµm particle sizes) containing mobile phase Buffer
and acetonitrile in ratio (65:35). The flow rate was set at 1.0 mL/minute and the injection volume was 10Î¼L. The run time of
injection is near about 15 min. In this method use Waters HPLC system with Reflective Index detector. The linearity of method
was 50 % to 100 %, the correlation coefficient was found to be 0.9999. Eluents were monitored by reflective index detector. The
liquid chromatographic method was validated with respect to specificity, precision (% RSD about 0.71%). Stability an analytical
solution (% deviation from initial area count found in limit Â±2%)</Abstract><Email>pkpatelyug@yahoo.com</Email><articletype>Review</articletype><volume>4</volume><issue>1</issue><year>2014</year><keyword>Memantine hydrochloride, Working Standard and sample,HPLC with reflective index detector</keyword><AUTHORS>Ajay Kumar Gupta,PK Patel,JJ Vora,Shivani Patel</AUTHORS><afflication>Department of Chemistry, Shri JJTU University, Rajasthan, India, Department of Chemistry, M.G Sciences Institute, Ahmedabad, Gujarat, India., Department of Chemistry, M.G Sciences Institute, Ahmedabad, Gujarat, India.,Department of Pharmaceutical Chemistry, L.M College of Pharmacy, Ahmedabad, Gujarat, India</afflication></Article></Articles>